Benchmark IMRT evaluation of a Co-60 MRI-guided radiation therapy system

H. Omar Wooten, Vivian Rodriguez, Olga Green, Rojano Kashani, Lakshmi Santanam, Kari Tanderup, Sasa Mutic, H. Harold Li

Research output: Contribution to journalArticlepeer-review

47 Scopus citations

Abstract

A device for MRI-guided radiation therapy (MR-IGRT) that uses cobalt-60 sources to deliver intensity modulated radiation therapy is now commercially available. We investigated the performance of the treatment planning and delivery system against the benchmark recommended by the American Association of Physicists in Medicine (AAPM) Task Group 119 for IMRT commissioning and demonstrated that the device plans and delivers IMRT treatments within recommended confidence limits and with similar accuracy as linac IMRT.

Original languageEnglish
Pages (from-to)402-405
Number of pages4
JournalRadiotherapy and Oncology
Volume114
Issue number3
DOIs
StatePublished - Mar 1 2015

Keywords

  • Co-60
  • IMRT
  • MRI-Guided radiation therapy

Fingerprint

Dive into the research topics of 'Benchmark IMRT evaluation of a Co-60 MRI-guided radiation therapy system'. Together they form a unique fingerprint.

Cite this