T1 - Assessment of intrathoracic impedance algorithm in the pediatric and adult congenital population
AU - Silva, Jennifer N.A.
AU - Von Bergen, Nicholas H.
AU - Dubin, Anne M.
AU - Collins, Kathryn K.
AU - Chen, Jiajing
AU - Bowman, Tammy M.
AU - Van Hare, George F.
N1 - Publisher Copyright:
©2014 Wiley Periodicals, Inc.
PY - 2014/9/1
Y1 - 2014/9/1
N2 - Background Decreased intrathoracic impedance has been used in adults to predict heart failure (HF) exacerbations. A commercial algorithm, OptiVol® (Medtronic Inc., Minneapolis, MN, USA), identifies patients with decreased impedance. We sought to determine the specificity, sensitivity, and positive predictive value (PPV) of OptiVol for predicting HF exacerbation or increased arrhythmia burden in pediatric and adult congenital heart disease (CHD) patients.Methods A multicenter retrospective chart review was undertaken. Inclusion criteria were: (1) <19 years or CHD adults, (2) an implanted device with OptiVol capability, (3) implanted between April 9 and September 6, and (4) follow-up of >30 days postimplant. Clinical events were defined as clinical HF exacerbation/hospital admission, initiation/uptitration of medication, or increased arrhythmia burden.Results Seventy-two patients (19 ± 9 years) were identified with the following indications: 20% dilated cardiomyopathy (DCM), 11% hypertrophic cardiomyopathy (HCM), 43% CHD, 15% channelopathy, and 11% other. Thirty-nine had 122 OptiVol crossings (median 2, range 1-11); 30% were linked to a cause. The remaining 33 had no crossing, though 17 had 89 clinical events. The clinical event rate was 19% greater in patients with crossings, though not statistically significant (P = 0.4). The algorithm had a 59% sensitivity, 52% specificity, and 62% PPV. Clinical HF exacerbation and arrhythmia burden did not significantly correlate with decreased impedance though uptitration or initiation of HF medication did correlate significantly (P = 0.03).Conclusion The algorithm sensitivity for pediatric DCM, HCM, CHD, and adult CHD was equivalent to the general adult population. Further studies are warranted to assess whether inaccuracy in prediction is secondary to the algorithm or to differences in the clinical response of pediatric/CHD patients.
AB - Background Decreased intrathoracic impedance has been used in adults to predict heart failure (HF) exacerbations. A commercial algorithm, OptiVol® (Medtronic Inc., Minneapolis, MN, USA), identifies patients with decreased impedance. We sought to determine the specificity, sensitivity, and positive predictive value (PPV) of OptiVol for predicting HF exacerbation or increased arrhythmia burden in pediatric and adult congenital heart disease (CHD) patients.Methods A multicenter retrospective chart review was undertaken. Inclusion criteria were: (1) <19 years or CHD adults, (2) an implanted device with OptiVol capability, (3) implanted between April 9 and September 6, and (4) follow-up of >30 days postimplant. Clinical events were defined as clinical HF exacerbation/hospital admission, initiation/uptitration of medication, or increased arrhythmia burden.Results Seventy-two patients (19 ± 9 years) were identified with the following indications: 20% dilated cardiomyopathy (DCM), 11% hypertrophic cardiomyopathy (HCM), 43% CHD, 15% channelopathy, and 11% other. Thirty-nine had 122 OptiVol crossings (median 2, range 1-11); 30% were linked to a cause. The remaining 33 had no crossing, though 17 had 89 clinical events. The clinical event rate was 19% greater in patients with crossings, though not statistically significant (P = 0.4). The algorithm had a 59% sensitivity, 52% specificity, and 62% PPV. Clinical HF exacerbation and arrhythmia burden did not significantly correlate with decreased impedance though uptitration or initiation of HF medication did correlate significantly (P = 0.03).Conclusion The algorithm sensitivity for pediatric DCM, HCM, CHD, and adult CHD was equivalent to the general adult population. Further studies are warranted to assess whether inaccuracy in prediction is secondary to the algorithm or to differences in the clinical response of pediatric/CHD patients.
KW - device diagnostics
KW - heart failure exacerbation
KW - intrathoracic impedance
UR - http://www.scopus.com/inward/record.url?scp=84908252905&partnerID=8YFLogxK
U2 - 10.1111/pace.12372
DO - 10.1111/pace.12372
M3 - Article
C2 - 24646399
AN - SCOPUS:84908252905
SN - 0147-8389
VL - 37
SP - 1174
EP - 1180
JO - Pacing and Clinical Electrophysiology
JF - Pacing and Clinical Electrophysiology
IS - 9
ER -