T1 - Adaptation of the emit theophylline assay to kinetic analyzers
T2 - The relationship of reaction kinetics to calculation procedures
AU - Dietzler, David N.
AU - Weidner, Noel
AU - Tieber, Virginia L.
AU - McDonald, Jay M.
AU - Smith, Carl H.
AU - Ladenson, Jack H.
AU - Leckie, Mary P.
PY - 1980/2/28
Y1 - 1980/2/28
N2 - We investigated the kinetic characteristics of an enzyme immunoassay system (EMIT) for the determination of theophylline. Less than 10% of the glucose-6-phosphate and NAD+ are consumed and the glucose-6-P-dehydrogenase-theophylline complex is essentially saturated with these substrates during the course of the reaction. However, apparently as the result of antibody heterogeneity, the rate does change during the course of the reaction. As a result, the values of the kinetic parameters for the theophylline saturation curve vary with the timing interval chosen for measurement. We show that the values of these parameters govern the application of the available methods of calculating EMIT data, the graphical procedure suggested by the manufacturer, logit-log, log-log, and curve fitting. These studies (1) explain the basis of the graphical procedure and why it does not always provide proper calibration, (2) show that to appropriately understand the application of any of the calculation procedures one should determine the values of the kinetic constants of the theophylline saturation curve in the particular assay conditions, and (3) illustrate simple, practical procedures for determining the values of these constants for any instrument-EMIT assay system. Specific illustrations are shown for the EMIT theophylline system with two kinetic analyzers, the Abbott ABA-100 and Gilford 3500, with which markedly different reaction conditions are used.
AB - We investigated the kinetic characteristics of an enzyme immunoassay system (EMIT) for the determination of theophylline. Less than 10% of the glucose-6-phosphate and NAD+ are consumed and the glucose-6-P-dehydrogenase-theophylline complex is essentially saturated with these substrates during the course of the reaction. However, apparently as the result of antibody heterogeneity, the rate does change during the course of the reaction. As a result, the values of the kinetic parameters for the theophylline saturation curve vary with the timing interval chosen for measurement. We show that the values of these parameters govern the application of the available methods of calculating EMIT data, the graphical procedure suggested by the manufacturer, logit-log, log-log, and curve fitting. These studies (1) explain the basis of the graphical procedure and why it does not always provide proper calibration, (2) show that to appropriately understand the application of any of the calculation procedures one should determine the values of the kinetic constants of the theophylline saturation curve in the particular assay conditions, and (3) illustrate simple, practical procedures for determining the values of these constants for any instrument-EMIT assay system. Specific illustrations are shown for the EMIT theophylline system with two kinetic analyzers, the Abbott ABA-100 and Gilford 3500, with which markedly different reaction conditions are used.
UR - https://www.scopus.com/pages/publications/0018842483
U2 - 10.1016/0009-8981(80)90242-9
DO - 10.1016/0009-8981(80)90242-9
M3 - Article
C2 - 6987012
AN - SCOPUS:0018842483
SN - 0009-8981
VL - 101
SP - 163
EP - 181
JO - Clinica Chimica Acta
JF - Clinica Chimica Acta
IS - 2-3
ER -