TY - JOUR
T1 - Accuracy and consistency of respiratory gating in abdominal cancer patients
AU - Ge, Jiajia
AU - Santanam, Lakshmi
AU - Yang, Deshan
AU - Parikh, Parag J.
PY - 2013/3/1
Y1 - 2013/3/1
N2 - Purpose: To evaluate respiratory gating accuracy and intrafractional consistency for abdominal cancer patients treated with respiratory gated treatment on a regular linear accelerator system. Methods and Materials: Twelve abdominal patients implanted with fiducials were treated with amplitude-based respiratory-gated radiation therapy. On the basis of daily orthogonal fluoroscopy, the operator readjusted the couch position and gating window such that the fiducial was within a setup margin (fiducial-planning target volume [f-PTV]) when RPM indicated "beam-ON." Fifty-five pre- and post-treatment fluoroscopic movie pairs with synchronized respiratory gating signal were recorded. Fiducial motion traces were extracted from the fluoroscopic movies using a template matching algorithm and correlated with f-PTV by registering the digitally reconstructed radiographs with the fluoroscopic movies. Treatment was determined to be "accurate" if 50% of the fiducial area stayed within f-PTV while beam-ON. For movie pairs that lost gating accuracy, a MATLAB program was used to assess whether the gating window was optimized, the external-internal correlation (EIC) changed, or the patient moved between movies. A series of safety margins from 0.5 mm to 3 mm was added to f-PTV for reassessing gating accuracy. Results: A decrease in gating accuracy was observed in 44% of movie pairs from daily fluoroscopic movies of 12 abdominal patients. Three main causes for inaccurate gating were identified as change of global EIC over time (∼43%), suboptimal gating setup (∼37%), and imperfect EIC within movie (∼13%). Conclusions: Inconsistent respiratory gating accuracy may occur within 1 treatment session even with a daily adjusted gating window. To improve or maintain gating accuracy during treatment, we suggest using at least a 2.5-mm safety margin to account for gating and setup uncertainties.
AB - Purpose: To evaluate respiratory gating accuracy and intrafractional consistency for abdominal cancer patients treated with respiratory gated treatment on a regular linear accelerator system. Methods and Materials: Twelve abdominal patients implanted with fiducials were treated with amplitude-based respiratory-gated radiation therapy. On the basis of daily orthogonal fluoroscopy, the operator readjusted the couch position and gating window such that the fiducial was within a setup margin (fiducial-planning target volume [f-PTV]) when RPM indicated "beam-ON." Fifty-five pre- and post-treatment fluoroscopic movie pairs with synchronized respiratory gating signal were recorded. Fiducial motion traces were extracted from the fluoroscopic movies using a template matching algorithm and correlated with f-PTV by registering the digitally reconstructed radiographs with the fluoroscopic movies. Treatment was determined to be "accurate" if 50% of the fiducial area stayed within f-PTV while beam-ON. For movie pairs that lost gating accuracy, a MATLAB program was used to assess whether the gating window was optimized, the external-internal correlation (EIC) changed, or the patient moved between movies. A series of safety margins from 0.5 mm to 3 mm was added to f-PTV for reassessing gating accuracy. Results: A decrease in gating accuracy was observed in 44% of movie pairs from daily fluoroscopic movies of 12 abdominal patients. Three main causes for inaccurate gating were identified as change of global EIC over time (∼43%), suboptimal gating setup (∼37%), and imperfect EIC within movie (∼13%). Conclusions: Inconsistent respiratory gating accuracy may occur within 1 treatment session even with a daily adjusted gating window. To improve or maintain gating accuracy during treatment, we suggest using at least a 2.5-mm safety margin to account for gating and setup uncertainties.
UR - http://www.scopus.com/inward/record.url?scp=84873279303&partnerID=8YFLogxK
U2 - 10.1016/j.ijrobp.2012.05.006
DO - 10.1016/j.ijrobp.2012.05.006
M3 - Article
C2 - 22717241
AN - SCOPUS:84873279303
SN - 0360-3016
VL - 85
SP - 854
EP - 861
JO - International Journal of Radiation Oncology Biology Physics
JF - International Journal of Radiation Oncology Biology Physics
IS - 3
ER -