T1 - A method for visualization of ventricular fibrillation
T2 - Design of a cooled fiberoptically coupled image intensified CCD data acquisition system incorporating wavelet shrinkage based adaptive filtering
AU - Witkowski, Francis X.
AU - Leon, L. Joshua
AU - Penkoske, Patricia A.
AU - Clark, Robert B.
AU - Spano, Mark L.
AU - Ditto, William L.
AU - Giles, Wayne R.
PY - 1998/3
Y1 - 1998/3
N2 - The measurement of cardiac transmembrane potential changes with voltage sensitive dyes is in increasing use. Detection of these very small fluorescent alterations using large multiplexed arrays, such as charge coupled device (CCD) cameras at high sampling rates, has proven challenging and usually requires significant averaging to improve the signal-to-noise ratio. To minimize the damage of living tissue stained with voltage sensitive dyes, excitation photon exposure must be limited, with the inevitable consequence of diminishing the fluorescence that is generated. State-of-the-art high frame rate CCD cameras have read noise levels in the 5-10 e- rms range, which is at least two orders of magnitude above that required to detect voltage sensitive dye alterations at individual pixels corresponding to 1 mm2 heart regions illuminated with levels of 100 mW/cm2 at frame rates approaching 1000 frames/sec. Image intensification is thus required prior to photon quantification. We report here the development of such a data acquisition system using commercially available hardware. Additionally, in the past ten years, a mathematical theory of multiresolution has been developed, and new building blocks called wavelets, allow a signal to be observed at different resolutions. Wavelet analysis also makes possible a new method of extricating signals from noise. We have incorporated spatially adaptive filters based on wavelet denoising of individual pixels to significantly reduce the multiple noise sources present in the acquired data.
AB - The measurement of cardiac transmembrane potential changes with voltage sensitive dyes is in increasing use. Detection of these very small fluorescent alterations using large multiplexed arrays, such as charge coupled device (CCD) cameras at high sampling rates, has proven challenging and usually requires significant averaging to improve the signal-to-noise ratio. To minimize the damage of living tissue stained with voltage sensitive dyes, excitation photon exposure must be limited, with the inevitable consequence of diminishing the fluorescence that is generated. State-of-the-art high frame rate CCD cameras have read noise levels in the 5-10 e- rms range, which is at least two orders of magnitude above that required to detect voltage sensitive dye alterations at individual pixels corresponding to 1 mm2 heart regions illuminated with levels of 100 mW/cm2 at frame rates approaching 1000 frames/sec. Image intensification is thus required prior to photon quantification. We report here the development of such a data acquisition system using commercially available hardware. Additionally, in the past ten years, a mathematical theory of multiresolution has been developed, and new building blocks called wavelets, allow a signal to be observed at different resolutions. Wavelet analysis also makes possible a new method of extricating signals from noise. We have incorporated spatially adaptive filters based on wavelet denoising of individual pixels to significantly reduce the multiple noise sources present in the acquired data.
UR - http://www.scopus.com/inward/record.url?scp=0346207529&partnerID=8YFLogxK
U2 - 10.1063/1.166290
DO - 10.1063/1.166290
M3 - Article
C2 - 12779713
AN - SCOPUS:0346207529
SN - 1054-1500
VL - 8
SP - 94
EP - 102
JO - Chaos
JF - Chaos
IS - 1
ER -