A device for applying static loads to prosthetic limbs of transtibial amputees during spiral CT examination

M. T. Madsen, J. Haller, P. K. Commean, M. W. Vannier

Research output: Contribution to journalArticlepeer-review

19 Scopus citations

Abstract

This paper describes a device that allows the imaging of prosthetic legs under load conditions using spiral computerized tomography (CT). The device consists of a chair and a vertical footplate mounted on a 17.8 × 1.9 cm oak board that is 197 cm long. The load device can be easily positioned onto the CT bed. A subject sits in the chair and applies force by pushing the foot portion of the prosthesis against the footplate. The magnitude of the force is monitored by a digital force gauge coupled to the footplate. Because the load is borne by the hips and lower back of the subject against the chair, substantial forces can be generated and steadily maintained for the 20-45 s duration of the CT study. This device has been used successfully with 19 transtibial amputees, allowing the acquisition of spiral CT studies with half and full body weight loads.

Original languageEnglish
Pages (from-to)383-387
Number of pages5
JournalJournal of Rehabilitation Research and Development
Volume37
Issue number4
StatePublished - 2000

Keywords

  • Loading device
  • Prosthetic fit
  • Spiral computerized tomography (CT)
  • Transtibial amputation

Fingerprint

Dive into the research topics of 'A device for applying static loads to prosthetic limbs of transtibial amputees during spiral CT examination'. Together they form a unique fingerprint.